HI83300 Series Multiparameter Photometers with Digital pH Electrode Input

HI83300 is a compact, multiparameter photometer for use in the lab or in the field. The meter is one of the most advanced photometers available with a innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. This meter has 60 different programmed methods measuring 37 key water quality parameters and also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves.

To save valuable laboratory benchtop space, the HI83300 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.

Advanced optical system

Unparalleled performance from a benchtop photometer

Digital pH electrode input

Save valuable bench space with one meter that works both as a photometer and as a laboratory pH meter

Absorbance measuring mode

Allows performance verification using CAL Check standards

Κωδικός προϊόντος: HI83300 Κατηγορίες: , Ετικέτα:

Πληροφορίες προϊόντος

From aluminum to zinc, the HI83300 benchtop photometer measures 37 different key water quality parameters using 60 different methods. This photometer features an innovative optical system that uses LEDs, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.

A digital pH electrode input allows the user to measure pH by a traditional glass electrode. The digital pH electrode has a built in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.

The HI83300 offers an absorbance measuring mode that allows for CAL Check standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the five wavelengths of light (420 nm, 466 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for educators to teach the concept of absorbance by using the Beer-Lambert Law.

Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.

 

Features/Benefits:

Backlit 128 x 64 Pixel Graphic LCD Display

  • Backlit graphic display allows for easy viewing in low light conditions
  • The 128 x 64 Pixel LCD allows for a simplified user interface with virtual keys  and on-screen help to guide the user through use of the meter

Built-in Reaction Timer for Photometric Measurements

  • The measurement is taken after the countdown timer expires.
  • Countdown timer ensures that all readings are taken at the appropriate reaction intervals regardless of user for better consistency in measurements

Absorbance mode

  • Hanna’s exclusive CAL Check cuvettes for validation of light source and detector
  • Allows for the user to plot concentration versus absorbance for a specific wavelength for use with user supplied chemistry or for teaching principles of photometry

Units of Measure

  • Appropriate unit of measure along with chemical form is displayed along with reading

Result Conversion

  • Automatically convert readings to other chemical forms with the touch of a button

Cuvette Cover

  • Aids in preventing stray light from affecting measurements

Digital pH Electrode Input

  • Measure pH and temperature with a single probe
  • Good Laboratory Practice (GLP) to track calibration information including date, time, buffers used, offset and slope for traceability
  • pH CAL Check alerts user to potential problems during the calibration process
  • Space saving having a pH meter and photometer built into one meter

Data Logging

  • Up to 1000 photometric and pH readings can be stored by simply pressing the dedicated LOG button. Logged readings are just as easily recalled by pressing the RCL button
  • Sample ID and User ID information can be added to a logged reading using alphanumeric keypad

Connectivity

  • Logged readings can be quickly and easily transferred to a flash drive using the USB-A host port or to a computer using the micro USB-B port
  • Data is exported as a .CSV file for use with common spreadsheet programs

Battery Status Indicator

  • Indicates the amount of battery life left

Error Messages

  • Photometric error messages include no cap, high zero, and standard too low
  • pH calibration messages include clean electrode, check buffer and check probe

On-screen Features

Method Selection
Users can easily select any one of the 60 measurement methods via the dedicated METHOD button.

 

 

 

Data Logging
Up to 1000 measurement readings can be logged with user and sample ID and recalled for future use.

 

 

 

pH Measurement Mode
Selecting the pH measurement mode allows for the photometer to be used as a professional pH meter with many features including temperature compensated measurements, automatic two point calibration, and GLP.

 

 

Advanced Optical System

HI83300 is designed with an innovative optical system that incorporates a beam splitter so that light can be used for absorbance readings and for a reference detector. The reference detector monitors the intensity of light and modulates when there is drift due to power fluctuation or the heating of the optical components. Each part has an important role in providing unparalleled performance from a photometer.

 

 

Specifications

SKUHI83300-02
Product NameMultiparameter photometer, with digital pH (230V)
Stock Status MessageCall if accurate shipping time is required
Quote RequiredNo
pH RangePhotometer
pH ResolutionPhotometer
pH AccuracyPhotometer
pH CalibrationAutomatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01)
pH Temperature CompensationAutomatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used
pH CAL Check (electrode diagnostics)clean electrode and check buffer/check probe displayed during calibration
pH MethodPhotometer
pH-mV Range±1000 mV
pH-mV Resolution0.1 mV
pH-mV Accuracy±0.2 mV
Dissolved Oxygen Range0.0 to 10.0 mg/L (as O2)
Dissolved Oxygen Resolution0.1 mg/L
Dissolved Oxygen Accuracy±0.4 mg/L ±3% of reading
Dissolved Oxygen Measurement MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method
Absorbance Range0.000 to 4.000 Abs
Absorbance Resolution0.001 Abs
Absorbance Accuracy+/-0.003Abs @ 1.000 Abs
Alkalinity RangeFreshwater
Alkalinity Resolution1 mg/L
Alkalinity Accuracy±5 mg/L ±5% of reading
Alkalinity MethodColorimetric method
Aluminum Range0.00 to 1.00 mg/L (as Al3+)
Aluminum Resolution0.01 mg/L
Aluminum Accuracy±0.04 mg/L ±4% of reading
Aluminum MethodAdaptation of the aluminon method
Ammonia RangeLow Range
Ammonia Resolution0.01 mg/L; 0.1 mg/L
Ammonia AccuracyLow Range
Ammonia MethodAdaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method
Anionic Surfactants Range0.00 to 3.50 mg/L (as SDBS)
Anionic Surfactants Resolution0.01 mg/L
Anionic Surfactants Accuracy±0.04 mg/L ±3% of reading
Anionic Surfactants MethodAdaptation of the USEPA method 425.1 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 5540C, Anionic Surfactants as MBAS.
Bromine Range0.00 to 8.00 mg/L (as Br2)
Bromine Resolution0.01 mg/L
Bromine Accuracy±0.08 mg/L ±3% of reading
Bromine MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method.
Calcium RangeFreshwater
Calcium Resolution1 mg/L
Calcium AccuracyFreshwater
Calcium MethodFreshwater
Chloride Range0.0 to 20.0 mg/L (as Cl⁻)
Chloride Resolution0.1 mg/L
Chloride Accuracy±0.5 mg/L ±6% of reading at 25 °C
Chlorine Dioxide Range0.00 to 2.00 mg/L (as ClO2)
Chlorine Dioxide Resolution0.01 mg/L
Chlorine Dioxide Accuracy±0.10 mg/L ±5% of reading
Chlorine Dioxide MethodAdaptation of the Chlorophenol Red method.
Free Chlorine Range0.00 to 5.00 mg/L (as Cl2)
Ultra Low Range
Free Chlorine Resolution0.01 mg/L
Ultra Low Range
Free Chlorine Accuracy±0.03 mg/L ±3% of reading
Ultra Low Range
Total Chlorine Range0.00 to 5.00 mg/L (as Cl2)
Ultra Low Range
Total Chlorine Resolution0.01 mg/L
Ultra Low Range
Total Chlorine Accuracy±0.03 mg/L ±3% of reading
Ultra Low Range
Chlorine MethodAdaptation of the EPA 330.5 DPD method
Free Chlorine (ULR) & Total Chlorine (UHR)
Chromium, Hexavalent RangeLow Range
Chromium, Hexavalent Resolution1 μg/L
Chromium, Hexavalent AccuracyLow Range
Chromium, Hexavalent MethodAdaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method.
Color, Water Range0 to 500 PCU (Platinum Cobalt Units)
Color, Water Resolution1 PCU
Color, Water Accuracy±10 PCU ±5% of reading
Color, Water MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method.
Copper RangeLow Range
Copper Resolution0.001 mg/L; 0.01 mg/L
Copper AccuracyLow Range
Copper MethodAdaptation of the EPA bicinchoninate method
Cyanuric Acid Range0 to 80 mg/L (as CYA)
Cyanuric Acid Resolution1 mg/L
Cyanuric Acid Accuracy±1 mg/L ±15% of reading
Cyanuric Acid MethodAdaptation of the turbidimetric method
Fluoride RangeLow Range
Fluoride Resolution0.01 mg/L; 0.1 mg/L
Fluoride AccuracyLow Range
Fluoride MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, SPADNS method
Hardness, Total RangeLow Range
Hardness, Total Resolution1 mg/L
Hardness, Total AccuracyLow Range
Hardness, Total MethodAdaptation of the EPA recommended method 130.1
Hardness, Calcium Range0.00 to 2.70 mg/L (as CaCO3)
Hardness, Calcium Resolution0.01 mg/L
Hardness, Calcium Accuracy±0.11 mg/L ±5% of reading
Hardness, Calcium MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method
Hardness, Magnesium Range0.00 to 2.00 mg/L (as CaCO3)
Hardness, Magnesium Resolution0.01 mg/L
Hardness, Magnesium Accuracy±0.11 mg/L ±5% of reading
Hardness, Magnesium MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA Colorimetric method
Hydrazine Range0 to 400 μg/L (as N2H4)
Hydrazine Resolution1 μg/L
Hydrazine Accuracy±4% of full scale reading
Hydrazine MethodAdaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method
Iodine Range0.0 to 12.5 mg/L (as I2)
Iodine Resolution0.1 mg/L
Iodine Accuracy±0.1 mg/L ±5% of reading
Iodine MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method
Iron RangeLow Range
Iron Resolution0.001 mg/L; 0.01 mg/L
Iron AccuracyLow Range
Iron MethodLow Range
Magnesium Range0 to 150 mg/L (as Mg2+)
Magnesium Resolution1 mg/L
Magnesium Accuracy±5 mg/L ±3% of reading
Magnesium MethodAdaptation of the Calmagite method
Manganese RangeLow Range
Manganese Resolution1 μg/L; 0.1 mg/L
Manganese AccuracyLow Range
Manganese MethodLow Range
Molybdenum Range0.0 to 40.0 mg/L (as Mo6+)
Molybdenum Resolution0.1 mg/L
Molybdenum Accuracy±0.3 mg/L ±5% of reading
Molybdenum MethodAdaptation of the mercaptoacetic acid method
Nickel RangeLow Range
Nickel Resolution0.001 mg/L; 0.01 g/L
Nickel AccuracyLow range
Nickel MethodLow Range
Nitrate Range0.0 to 30.0 mg/L (as NO3-- N)
Nitrate Resolution0.1 mg/L
Nitrate Accuracy±0.5 mg/L ±10% of reading
Nitrate MethodAdaptation of the cadmium reduction method
Nitrite RangeFreshwater
Low Range
Nitrite ResolutionFreshwater
Nitrite AccuracyFreshwater
Low Range
Nitrite MethodAdaptation of the EPA Diazotization method 354.1
Oxygen, Scavenger Range0 to 1000 μg/L (as DEHA)
0.00 to 1.50 mg/L (as Carbohydrazide)
0.00 to 2.50 mg/L (as Hydroquinone)
0.00 to 4.50 mg/L (as ISO-ascorbic acid)
Oxygen, Scavenger Resolution1 μg/L (DEHA); 0.01 mg/L
Oxygen, Scavenger Accuracy±5 μg/L ±5% of reading (DEHA)
Oxygen, Scavenger MethodAdaptation of the iron reduction method
Ozone Range0.00 to 2.00 mg/L (as O3)
Ozone Resolution0.01 mg/L
Ozone Accuracy±0.02 mg/L ±3% of reading
Ozone MethodColorimetric DPD Method
Phosphate RangeFreshwater
Low Range
Phosphate ResolutionFreshwater
Phosphate AccuracyFreshwater
Low Range
Phosphate MethodFreshwater Low Range
Potassium Range0.0 to 20.0 mg/L (as K)
Potassium Resolution0.1 mg/L
Potassium Accuracy±3.0 mg/L ±7% of reading
Potassium MethodAdaptation of the Turbidimetric Tetraphenylborate method
Silica RangeLow Range
Silica Resolution0.01 mg/L; 1 mg/L
Silica AccuracyLow Range
Silica MethodLow Range
Silver Range0.000 to 1.000 mg/L (as Ag)
Silver Resolution0.001 mg/L
Silver Accuracy±0.020 mg/L ±5% of reading
Silver MethodAdaptation of the PAN method
Sulfate Range0 to 150 mg/L (as SO42-)
Sulfate Resolution1 mg/L
Sulfate Accuracy±5 mg/L ±3% of reading
Sulfate MethodTurbidimetric - Sulfate is precipitated with barium chloride crystals
Zinc Range0.00 to 3.00 mg/L (as Zn)
Zinc Resolution0.01 mg/L
Zinc Accuracy±0.03 mg/L ±3% of reading
Zinc MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method
Input Channels1 pH electrode input and 5 photometer wavelengths
pH Electrodedigital pH electrode (not included)
Photometer/Colorimeter Light Source5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters
Photometer/Colorimeter Light Detectorsilicon photodetector
Bandpass Filter Bandwidth8 nm
Bandpass Filter Wavelength Accuracy±1 nm
Cuvette Typeround, 24.6 mm
Number of Methods128 max.
GLPcalibration data for connected pH electrode
Display128 x 64 pixel LCD with backlight
Logging Typelog on demand with user name and sample ID optional input
Logging Memory1000 readings
ConnectivityUSB-A host for flash drive; micro-USB-B for power and computer connectivity
Power Supply5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Battery Type/Life3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement
Environment0 to 50.0 oC (32 to 122.0 oF); 0 to 95% RH, non-condensing
Weight1.0 kg (2.2 lbs.)
Dimensions206 x 177 x 97 mm (8.1 x 7.0 x 3.8")
Ordering InformationHI83300 is supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual.