HI97716 Bromine Portable Photometer

354,00 

The HI97716 portable photometer is for the measurement of bromine. Hanna’s portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.

Stable Light Source:

The internal reference system of the HI-97716 photometer compensates for any drifts due to power fluctuations or ambient temperature changes. With a stable source of light the readings are fast and stable between your blank (zero) measurement and sample measurement.

High Efficiency Light Source:

LED light sources offer superior performance compared to tungsten lamps. LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce little heat, which could otherwise affect electronic stability.

High Quality Filters:

Improved optical filters ensure greater wavelength accuracy and allow a brighter, stronger signal to be received. The end result is higher measurement stability and less wavelength error.

Greater Light Yield:

A focusing lens collects all of the light that exits the cuvette, eliminating errors from imperfections and scratches that may be present in the glass. The use of the convex lens reduces the need for indexing cuvettes.

CAL Check Functionality:

Hanna’s exclusive CAL Check feature allows for performance verification and calibration of the meter using NIST traceable standards. Our CAL Check standard vials are developed to simulate a specific absorbance value at each wavelength to verify the accuracy of subsequent readings.

Κωδικός προϊόντος: HI97716 Κατηγορίες: , Ετικέτα:

Πληροφορίες προϊόντος

The HI97716 Bromine Portable Photometer combines accuracy and ease of use in an ergonomic, portable design. Using the ready-made reagents a user can accurately determine the concentration of bromine within a 0.00 to 10.00 mg/L (ppm) range.

The HI97716 offers many advanced features including the exclusive CAL Check function used for performance verification and calibration of the meter.

  • No warm up time before taking a measurement
  • Tutorial mode for step-by-step instructions
  • Uses either powder or cost saving liquid reagents
  • CAL Check

 

 

Specifications

Bromine Range0.00 to 10.00 mg/L (ppm)
Bromine Resolution0.01 mg/L (ppm)
Bromine Accuracy±0.08 mg/L ±3% of reading
Bromine Methodadaptation of the Standard Methods for the Examination of Water and Wastewater, 20th edition, DPD method
Photometer/Colorimeter Light Sourcetungsten lamp
Photometer/Colorimeter Light Detectorsilicon photocell with narrow band interference filter @ 525 nm
Battery Type/Life9V battery
Automatic Shut-Offafter ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder
Environment0 to 50°C (32 to 122°F); RH max 95% non-condensing
Weight360g (12.7oz.)
Dimensions193 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Ordering InformationHI97716 is supplied with sample cuvettes (2) with caps, 9V battery, instrument quality certificate and instruction manual. CAL Check™ standards and testing reagents sold separately.