    • Accepts pH electrode and EC probe signals simultaneously
    • Direct probe connection; Assures a positive electrical connection eliminating signal loss
    • Automatic Temperature Compensation (EC only)

    The HI 98143 series is designed to accept signals directly from a pH electrode and a conductivity probe at the same time.

Κωδικός προϊόντος: HI98143-04 Κατηγορίες: , Ετικέτα:

Πληροφορίες προϊόντος

  • Accepts pH electrode and EC probe signals simultaneously
  • Direct probe connection; Assures a positive electrical connection eliminating signal loss
  • Automatic Temperature Compensation (EC only)

The HI 98143 series is designed to accept signals directly from a pH electrode and a conductivity probe at the same time.

Direct connection of the probes to the transmitter assure a positive electrical connection with no signal loss. This transmitter is ideal for remote process control applications.

Four models are available, transmitting a 0-1 V, 0-4 V or 4-20 mA signal. The output signals are proportional to the input signals but independent of changes in load or cable capacitance. Compensation for the effects of temperature for EC measurements are performed by the transmitters’ Automatic Temperature Compenstion circuitry.

The transmitter can be connected to any pH or conductivity controller, recorder, PC or any data monitoring device that accepts 0 to 1 V, 0 to 4 V or 4 to 20 mA input. HI?98143 is an ideal tool for applications that require the monitoring of both pH and conductivity at the same time.

Order Information:
HI 98143-04 with 0-4 V isolated output, is supplied with instructions.

Specifications

Specifications
RangepH0 to 14 pH
EC0 to 10 mS/cm
Accuracy @ 20°CpH±0.5% f.s.
EC±2% f.s.
CalibrationpHmanual, 2 point, through offset and slope trimmers
ECmanual, 2 point, through 0 and 5 mS/cm trimmers
Temperatureautomatic, 0 to 60°C (32 to 132°F) with ß=2%/°C
Compensation (EC)
Output (isolated)0-4 V
pH ElectrodeHI 1001 pH electrode (suggested, not included), HI 1283 matching pin (not included)
EC ProbeHI 3001 (not included) with cell constant 2.1
CasingIP 54
Power Supply12-24 VDC
Environment0 to 50°C (32 to 122°F); RH max 95% non-condensing
Dimensions160 x 105 x 31 mm (6.3 x 4.1 x 1.2’’)