HI2216 pH/ISE/ORP/°C Benchtop Meter with 0.001 Resolution

HI2216 pH/ISE/ORP/°C Benchtop Meter with 0.001 Resolution

The HI2216 is an accurate and affordable pH/ISE/ORP/°C benchtop meter with 0.001 pH and ISE resolution. Up to a five-point pH calibration can be performed using seven standard buffers. In addition, two custom buffers can be used during calibration. Up to a two-point ISE calibration can be performed using five standard solutions.

All pH readings are automatically compensated for temperature variations from the supplied thermistor probe. The HI2216 provides GLP capabilities and memory for storage of up to 200 log-on-demand records 500 lot logging records.

 

 Click to download the product’s manual 

The HI2216 is an accurate and affordable pH/ISE/ORP/°C benchtop meter with 0.001 pH and ISE resolution. Up to a five-point pH calibration can be performed using seven standard buffers. In addition, two custom buffers can be used during calibration. Up to a two-point ISE calibration can be performed using five standard solutions. All pH readings are automatically compensated for temperature variations from the supplied thermistor probe. The HI2216 provides GLP capabilities and memory for storage of up to 200 log-on-demand records 500 lot logging records.

Details

The HI2216 is an affordable, high performing pH/ISE/ORP/°C benchtop meter. Operation is simplified to calibration, measurement range selection, and a memory recall of logged data.

To enter calibration mode the user simply presses the “CAL” button. The HI2216 can perform up to a 5-point pH calibration with 7 standard buffers; ISE calibration can be performed on up to 2 points with five standard solutions.

The HI2216 has two separate inputs for a BNC electrode and thermistor temperature probe. The BNC connector is a universal connector that is commonly available and allows the user to select from a wide variety of pH, ORP, or ISE electrodes for their particular application. The separate temperature probe provides for a high 0.2 oC accuracy measurement for precise temperature compensation.

The memory of the HI2216 allows for storage of up to 200 log-on-demand records and up to 500 lot logging records. Logged data, along with Good Laboratory Practice (GLP) data, can be recalled on the meter itself or transferred to a PC via USB cable.

Features at-a-glance

Simple User Interface – Operation is simple with limited features that only require the use of a couple of buttons. Readings are easy to view on the large, clear display.

Automatic Calibration – Automatic pH calibration can be performed at up to 5 points using 7 standard buffers (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, and 12.45) and 2 custom buffers. Automatic ISE calibration can be performed at up to 2 points using 5 standard buffers (0.1, 1, 10, 100, 1000 ppm).

HI1131B pH Electrode – The HI2216 is supplied with the HI1131B glass body, double junction, refillable pH electrode with an indicating sensor made of High Temperature (HT) glass. The double junction and HT glass design allows the HI1131B to be used in a wide variety of application from samples with metals and Tris buffer to samples at elevated temperatures.

Temperature Compensation – The HI2216offers automatic or manual temperature compensation of pH measurements over a wide range from -20.0 to 120.0°C.

mV mode – The HI2216 has a mV mode that can be used with a wide variety of ORP electrodes. The mV mode is also useful in checking the offset and slope of a pH or ISE electrode. Understanding the offset and slope characteristic is helpful in knowing if the probe is functioning correctly.

Stability Indicator – The HI2216 features a stability indicator. An hourglass appears on the display when there is instability in the reading. The hourglass disappears once the reading has stabilized. At that time a reading should be taken or stored.

Measurement Recall – The HI2216 has a “RCL” button to review log-on-demand, lot logging, or GLP information on the meter display.

Data Transfer – Data can be transferred to a PC with a USB cable and HI92000 software (both sold separately).

Specifications

pH Specifications 
pH Range–2.0 to 16.0 pH, -2.000 to 16.000 pH, -2.00 to 16.00 pH
pH Resolution0.1; 0.01; 0.001
pH Accuracy (@25C/77F)0.1 pH; 0.01 pH; 0.002 pH
pH Calibrationautomatic, up to five-point calibration with seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) and two custom buffers
pH Temperature Compensationautomatic or manual (with HI7662 temperature probe) from –20.0 to 120.0 °C (-4.0 to 248.0°F)
mV Range999.9 mV; 2000 mV
mV Resolution0.1 mV (999.9 mV); 1 mV (2000 mV)
mV Accuracy0.2 mV (999.9 mV); 1 mV (2000 mV)
Relative mV Offset Range2000 mV
Temperature Specifications
Temperature Range-20.0 to 120.0C (-4.0 to 248.0F)
Temperature Resolution0.1 C
Temperature Accuracy0.2 C
ISE Specifications
ISE Range0.001 to 19990 ppm
ISE Resolution0.001 ppm; 0.01 ppm; 0.1 ppm; 1 ppm; 10 ppm
ISE Accuracy @ 25 C/77 F0.5% FS
ISE Calibration Pointsautomatic, one or two points with five available buffers (0.1, 1, 10, 100, 1000 ppm)
Additional Specifications
Electrode/ProbeHI1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included)
Temperature ProbeHI7662 temperature probe, stainless steel with 1 m (3.3’) cable (included)
Logginglog-on-demand up to 200 records; autologging up to 500 records
Logging Interval5, 10, 30 seconds, 1, 2, 5, 10, 15, 30, 60, 120, 180 minutes stability logging (“StAb”)
Connectivityopto-isolated USB
Environment0 to 50C (32 to 122F; 273 to 323 K), RH max 95% non-condensing
Input Impedance10 Ohms
Power Supply12 VDC adapter (included)
Dimensions235 x 222 x 109 mm (9.2 x 8.7 x 4.3”)
Weight1.3 kg (2.9 lb)
Ordering InformationHI2216 is supplied with HI1131B pH electrode, HI7662 temperature probe, HI76404N electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI7082S 3.5 KCl electrolyte solution (30 mL), HI700601 cleaning solution sachet, 12 VDC adapter and instructions.